Other hardware may vary, but should always be consistent. Using an Isopharm pressure transducer and power supply, this should be around 1.1-1.2 volts (this will vary slightly depending on local atmospheric pressure and variations in the electronics).When you begin a calibration and TQSoft is waiting for stability at the low point, check the voltage being received while at atmospheric pressure before you begin using your pressure calibrator. There are a couple of quick checks you can do to ensure your calibration is going to plan…ġ. 300mBA is 0.75 volts, and at 3000mBA is 2.5 volts, it can then translate any further voltage received into a displayed pressure.įor this reason, during calibration, TQSoft will display a voltage rather than a pressure reading at the low and high points, then switch to displaying an accurate pressure value for the checkpoint. In other words, when TQSoft is told that the voltage created at e.g. Pressure calibration in TQSoft creates a straight-line relationship between the pressure reported by your pressure calibrator, and the voltage created by the pressure transducer.